A COUPLED LASER ABLATION PULSED FOURIER TRANSFORM MICROWAVE SPECTROMETER - INITIAL TESTS

Loading...
Thumbnail Image

Date

1989

Journal Title

Journal ISSN

Volume Title

Publisher

Ohio State University

Research Projects

Organizational Units

Journal Issue

Abstract

A pulsed Nd-YAG ablation source has been coupled to a Fabry-Perot cavity. Fourier transform microwave spectrometer. The pulsed laser is used to vaporize ``non-volatile'' substances in the throat of a pulsed nozzle source. The vaporized substances are entrained in an inert carrier gas and pulsed into the Fabry-Perot cavity of the Fourier transform spectrometer. Initial experiments have involved the first microwave measurement of the interstellar species $SiC_{2}$. The $1_{01}-0_{00}$ transition of $^{28}SiC_{2}$ was observed at 23600.242(4) MHz and the corresponding transitions of $^{29}Si$ and $^{30}Si$ were also observed. $SiC_{2}$ was vaporized from a silicon carbide target rod. The electric dipole moment has also been measured and found to be 2.393(6) D. In another experiment the $J = 3-2$ and 2-1 transitions of KCl have been observed and measured. In this case, liquid KCl was used to coat the SiC rod. Descriptions of these and subsequent experiments will be given.

Description

Author Institution: Molecular Spectroscopy Division, National Institute of Standards and Technology

Keywords

Citation