An optical system is described for introducing an auxiliary beam of light into the sample beam path of a Baird Associates double-beam infrared spectrophotometer. The auxiliary-beam intensity is a constant fraction of the reference-beam intensity. This arrangement avoids working in the nonlinear range of the reference-beam attenuator near complete closure. In addition, a dynamically balanced zero is obtained at complete sample absorption, a condition not satisfied by the instrument before the modification was introduced.

Description

Author Institution: The Dow Chemical Company