AN INEXPENSIVE GRATING SPECTROMETER CONVERSION FOR RAPID SCANNING, HIGH RESOLUTION WORK

Loading...
Thumbnail Image

Date

1960

Journal Title

Journal ISSN

Volume Title

Publisher

Ohio State University

Research Projects

Organizational Units

Journal Issue

Abstract

This paper describes the modification of a Perkin-Elmer Model 12-C spectrometer for higher resolution in the 2.5 to 5 micron region. Brackets attached to the optical bench permitted the spectrometer to be mounted vertically on the wall. A spiral Nichrome

Description

$^{1}$D. D. Erley and L. W. Hersher, The Dow Chemical Company, private communication. $^{2}$Merton-NPL diffraction grating, Ruling \#287A, Sir Howard Grubh, Parsons \& Co. Ltd., Newcastle-upon-Tyne, England. $^{3}$RP.3W thermocouple, Charles M. Reeder \& Co., Detroit, Michigan.
Author Institution: Commercial Solvents Corporation