Show simple item record

dc.creatorLogan, L. M.en_US
dc.creatorGowrinathan, S.en_US
dc.creatorLudwig, H. G.en_US
dc.date.accessioned2006-06-15T14:11:47Z
dc.date.available2006-06-15T14:11:47Z
dc.date.issued1979en_US
dc.identifier1979-Pi-04en_US
dc.identifier.urihttp://hdl.handle.net/1811/10836
dc.descriptionAuthor Institution:en_US
dc.description.abstractThe S/N capability of an idea lab Fourier Transform Spectrometer using a JgCdTe detector was improved by a factor of 30 to a level where system performance with a Nernst glower source was limited only by the extrinsic detector noise. The results reported here constitute a hardware demonstration of an analysis recently published by Zachor and Aaronson. Our analysis and hardware demonstation are in substantial agreement with the work of Zachor and Aaronson which was carried out independently. Diagnostic date will be presented which illustrate the distortion to the spectrum caused by errors in sampling. The characteristics/specifications for signal and sampling control electronics which provide dynamic compensation for sampling errors will be discussed, and test date showing the 30 fold improvement in S/N will be presented.en_US
dc.format.extent78925 bytes
dc.format.mimetypeimage/jpeg
dc.language.isoEnglishen_US
dc.publisherOhio State Universityen_US
dc.titleS/N ENHANCEMENT OF AN FTS SYSTEM BY DYNAMIC FILTER COMPENSATION FOR SAMPLING ERRORSen_US
dc.typearticleen_US