AUTOMATED WAVENUMBER DETERMINATION FOR DIODE LASER SPECTRA

Please use this identifier to cite or link to this item: http://hdl.handle.net/1811/13256

Show full item record

Files Size Format View
1994-TD-06.jpg 121.9Kb JPEG image Thumbnail of AUTOMATED WAVENUMBER DETERMINATION FOR DIODE LASER SPECTRA

Title: AUTOMATED WAVENUMBER DETERMINATION FOR DIODE LASER SPECTRA
Creators: Xu, Li-Hong; Lees, R. M.; Moruzzi, Giovanni
Issue Date: 1994
Abstract: An accurate and convenient two-channel wavenumber calibration program has developed for a Laser Photonics diode laser system. The method is based solely on comparison of sample and reference spectra, as recorded and stored by the standard sweep control and data acquisition facilities supplied with the system. The calibration program reads the sample and reference data files in Laser photonics format, extracts the spectral data, and converts them into standard IEEE data format. A simple procedure is implemented for peakfinding, with the sample peak wavenumbers calculated from a calibration curve determined by least squares fitting to the reference wavenumbers. Provided the reference spectrum has a suitable distribution of peaks, a calibration precision of $\pm 0.0002 cm^{-1}$ or better is obtained over the full width of the diode laser scan, of the same order as the original standard reference. Given the rapid ongoing improvement in availability of good reference spectra for all regions, this program offers a simple reliable and automatic procedure for accurate measurement of diode laser data.